|
Multi-channel fibre optic
dosimeter based on Optically Stimulated Luminescence for dose verification
during radiotherapy treatments
Within the WP
3.2, CEA LIST (Saclay) investigates OSL dosimetry for dose verification in
radiotherapy. It collaborates with the Institut Gustave Roussy
(Villejuif) on clinical validations and Laboratoire National Henri Becquerel
(LNHB, Saclay) on pre-clinical validation tests.
IGR is the
first European cancer treatment center. It provides a clinical expertise,
accelerator equipments as well as usual dosimetry measurements.
LNHB is a
certified laboratory for the measurement of dose. It provides the equipment and
methodology to calibrate the sensors with respect to absolute dose.
The quality
control in radiotherapy needs an efficient, reliable, easy to use and
cost-effective multi-sensor dosimeter system in order to check the conformity
between the prescribed doses and the absorbed doses delivered to the patient,
especially in the context of IMRT for which high dose gradients are observed.
The
measurement and data calculation/storage must be performed in a restricted time
(less than some minutes) in order to be compatible with a treatment protocol.
The sensors should be radiation transparent in order not to attenuate or scatter
the radiation. Finally, they should be calibrated for all the energies
used.
Since OSL
dosimeters are truly integrating dosimeters, they may be read after treatment.
They offer significant advantages over active dosimeters such as avoiding
Cerenkov and scintillation perturbations associated to a high accuracy in dose
measurement. However, these sensors may be also used as scintillators to measure
the dose rate. For most treatments, the absorbed dose may be measured after
irradiation (delayed measurement) but some treatments (such as Total Body
Irradiation - TBI) necessitates a real-time estimation of the cumulated dose in
order to stop the irradiation when it reaches the required level.
The OSL
technique conveys the additional advantage over TL of measurement in operation
(no need to collect the sensor) and over MOSFET, longer lifetime and lower
maintenance cost.
Optically
Stimulated Luminescence (OSL) dosimetry is similar to the well-known
thermoluminescence dosimetry (TL) except that the luminescence is stimulated by
light instead of heat, opening the way to a remote and operational optical
interrogation.
OSL dosimetry
with alumina conveys a lot of advantages (high sensitivity, very good linearity,
no fading at room temperature, low Z (~10)). The fiber sensors are compatible
with in-vivo applications (small (mm size) and sterilizable). They are
immune to electromagnetic perturbations and exhibit low energy and angular
dependences.
The OSL dosimeter system proposed by the CEA LIST operates in a continuous-wave
mode (CW-OSL) which is simple and convenient for performing delayed OSL
measurements. Moreover, the system incorporates an optical fibre switch (16
channels) (fig. 1-2) in a cost-effective approach.

Fig. 1:
View of the OSL dosimeter developed by the CEA LIST

Fig. 2:
Principle of the CEA LIST OSL dosimeter system
The
instrumentation, located in the irradiation room (protected by screens), is
handled from the control room by a laptop through a dedicated software written
in LabView.
The OSL
signals are read sequentially after treatment and integrated to provide the
doses. The light stimulation also fully bleaches the sensor crystals so that
they can be remotely read in operation (reusable for the next patient).
During Total
Body Irradiation (TBI) treatments, the optical switch is periodically rotated to
scan the sensors emissions and a routine calculates the cumulated dose for each
sensor (period ~ 2 seconds for 12 sensors) by integrating the radioluminescence
(RL) signals emitted by the sensors (in the absence of laser stimulation). Then,
the absorbed dose (i.e. at the end of the treatment) is measured by OSL
which also bleaches the crystal for a next usage. To compensate for the - albeit
relatively weak - stem effect perturbation (fibre Cerenkov and scintillation), a
second identical optical fibre (without detector) is placed in parallel with and
aside the first one. The luminescence from the reference fibre is then
subtracted from the RL signal measured with the sensor.
Each fibre
sensor consist of an optical cable with a SMA connector at one end and a molded
cylindrical head at the other (fig. 3), made of polymers for
radiation transparency. The a-Al2O3
crystal is 1 mm long and has a diameter of 1 mm. It is protected by a sleeve
rigidly fixed inside the cable and affixed near the end of a silica fibre (Æ =
600 µm, NA = 0.37). It is remotely stimulated via the optical fibre using
a doubled Nd-YAG laser (532 nm, 150 mW). The OSL is collected and guided back
along the same fibre to a photomultiplier tube through adequate optical filters.

Fig.
3:
OSL fibre sensors for external radiotherapy (Ø=6 mm)
Tests are
performed using an X-ray generator (80 keV, 0.15 Gy/min) and a Saturn 43 linear
accelerator (6 MV, 12 MV, 20 MV). The Fig. 4 shows a continuous irradiation by
the X-ray generator during 2 minutes. The RL (observed during the irradiation
exposure) provides an estimation of the dose rate. Then, the laser shutter is
opened to stimulate the OSL and the OSL signal (with a decaying shape) is
recorded and integrated (dose resolution ~ 1 mGy). The crystal is bleached
within 24 seconds (> 99.9 % of trap depletion) for 15 mW at the fibre end. An
additional time of about 6 s is needed to record the background noise that is
subtracted to provide the dose measurement.
OSL readings
may be programmed as the irradiation advances, following the planned treatment.

Fig. 4:
Typical RL/OSL signals obtained with the fibre sensors depicted on fig. 3
Pre-clinical
validation tests are planned on a Saturn 43 linear accelerator to check the
energy response (for several energies ranging from 1 MeV to 20 MeV), angular
response and dose reproducibility. Clinical validations will then be carried out
at IGR facilities to validate the instrumentation in real medical conditions on
body simulators (phantoms).
An industrial
transfer is planned for 2009 towards a french SME specialized in medical
dosimetry. However, other licenses are expected.
[1] O. Roy, S.
Magne, J.C. Gaucher, L. Albert, L. Dusseau, J.C. Bessière and P. Ferdinand, All
Optical Fiber Sensor based on Optically Stimulated Luminescence for Radiation
Detection, Optical Fiber Sensor Conference (OFS12), 1997, Williamsburg
(Virginia), USA
[2] S. Magne, P.
Ferdinand, Fiber Optic remote g dosimeters based on Optically Stimulated
Luminescence: State-of-the art at CEA, IRPA 11, Madrid, 23-27 May 2004
[3] G. Ranchoux,
S. Magne, J.P. Bouvet and P. Ferdinand, Fiber Remote Optoelectronic gamma
dosimetry based on Optically Stimulated Luminescence of Al2O3:C,
Rad. Prot. Dos. 100, 1-4 (2002) 255-260
[4] S. Magne, L. Auger, A.
Isambert, A. Bridier, P. Ferdinand, J. Barthe, Développement d’un dosimètre
multivoies à fibres optiques pour la radiothérapie, fondé sur la mesure de la
scintillation et de l’OSL de l’alumine, 45èmes journées Scientifiques
de la SFPM (Société Française de Physique Médicale), 7-9 juin 2006, Lyon (ENS)
[5] S. Magne, L.
Auger, A. Isambert, A. Bridier, P. Ferdinand, J. Barthe, Multi-channel fibre
optic dosimeter based on OSL for dose verification during radiotherapy
treatments, 18th Optical Fiber Sensor Conference, Cancun, Mexico,
October 23-27, 2006.
|
Home Page |
|